Adverse events in contemporary continuous-flow left ventricular assist devices: A multi-institutional comparison shows significant differences

J Thorac Cardiovasc Surg. 2016 Jan;151(1):177-89. doi: 10.1016/j.jtcvs.2015.09.100. Epub 2015 Nov 4.

Abstract

Objectives: We review differences in the incidence and timing of adverse events in patients implanted with continuous-flow left ventricular assist devices (LVADs), using the Mechanical Circulatory Support Research Network registry.

Methods: From May 2004 to September 2014, a total of 734 patients (591 men; median age: 59 years) underwent primary continuous-flow LVAD implantation at our institutions. Patients implanted with the HeartMate II (HMII) (560 [76%] patients), compared with the HeartWare ventricular assist device (HVAD; 174 [24%]) were more often receiving destination therapy (47% vs 20%; P < .01), had a lower preoperative creatinine level (1.2 vs 1.3; P = .01), and had less median preoperative right ventricular dysfunction (mild vs moderate; P < .01). Ischemic etiology, prior sternotomy, and median INTERMACS profile were similar.

Results: Overall mortality was 54 of 734 (7.4%); 41 of 560 (7.3%) in the HMII group, and 13 of 174 (7.5%) in the HVAD group (P = .95). Follow-up was available in 100% of early survivors for a median of 1 year (max: 10 years) and a total of 1120 patient-years of support (HMII: 940 patient-years [median: 1.1 years, max: 5.3 years] and HVAD: 180 patient-years [median: 0.6 year, max: 10.4 years]). On multivariable analysis, GI bleeding (P = .63), any infection (P = .32), driveline infection (P = .10), and pump thrombus (P = .64) were similar between devices while HeartWare HVAD was associated with higher risk of stroke (HR: 1.8, [1.25, 2.5], P = .003).

Conclusions: In this pooled analysis, a trend was found for higher incidence of percutaneous driveline infections in patients treated with the HMII; a higher incidence of stroke and time-related cumulative risk of any infection and stroke was found in patients treated with the HVAD, which was independently associated with higher stroke risk.

Keywords: heart failure; mechanical circulatory support; ventricular assist device.

Publication types

  • Comparative Study
  • Multicenter Study

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Chi-Square Distribution
  • Female
  • Gastrointestinal Hemorrhage / epidemiology
  • Heart Failure / diagnosis
  • Heart Failure / mortality
  • Heart Failure / physiopathology
  • Heart Failure / therapy*
  • Heart-Assist Devices / adverse effects*
  • Humans
  • Incidence
  • Kaplan-Meier Estimate
  • Male
  • Middle Aged
  • Multivariate Analysis
  • Proportional Hazards Models
  • Prosthesis Design
  • Prosthesis-Related Infections / epidemiology
  • Prosthesis-Related Infections / microbiology
  • Registries
  • Retrospective Studies
  • Risk Assessment
  • Risk Factors
  • Stroke / epidemiology
  • Thrombosis / epidemiology
  • Time Factors
  • Treatment Outcome
  • United States / epidemiology
  • Ventricular Function, Left
  • Ventricular Function, Right