Adaptive statistical iterative reconstruction reduces patient radiation dose in neuroradiology CT studies

  • Diagnostic Neuroradiology
  • Published:
Neuroradiology Aims and scope Submit manuscript

Abstract

Introduction

Adaptive statistical iterative reconstruction (ASIR) can decrease image noise, thereby generating CT images of comparable diagnostic quality with less radiation. The purpose of this study is to quantify the effect of systematic use of ASIR versus filtered back projection (FBP) for neuroradiology CT protocols on patients’ radiation dose and image quality.

Methods

We evaluated the effect of ASIR on six types of neuroradiologic CT studies: adult and pediatric unenhanced head CT, adult cervical spine CT, adult cervical and intracranial CT angiography, adult soft tissue neck CT with contrast, and adult lumbar spine CT. For each type of CT study, two groups of 100 consecutive studies were retrospectively reviewed: 100 studies performed with FBP and 100 studies performed with ASIR/FBP blending factor of 40 %/60 % with appropriate noise indices. The weighted volume CT dose index (CTDIvol), dose–length product (DLP) and noise were recorded. Each study was also reviewed for image quality by two reviewers. Continuous and categorical variables were compared by t test and free permutation test, respectively.

Results

For adult unenhanced brain CT, CT cervical myelography, cervical and intracranial CT angiography and lumbar spine CT both CTDIvol and DLP were lowered by up to 10.9 % (p < 0.001), 17.9 % (p = 0.005), 20.9 % (p < 0.001), and 21.7 % (p = 0.001), respectively, by using ASIR compared with FBP alone. Image quality and noise were similar for both FBP and ASIR.

Conclusion

We recommend routine use of iterative reconstruction for neuroradiology CT examinations because this approach affords a significant dose reduction while preserving image quality.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1

Similar content being viewed by others

Abbreviations

ASIR:

Adaptive statistical iterative reconstruction

CTDIvol :

Weighted volume CT dose index

DLP:

Dose–length product

FBP:

Filtered back projection

NI:

Noise index

References

  1. Pearce MS, Salotti JA, Little MP, McHugh K, Lee C, Kim KP et al (2012) Radiation exposure from ct scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet 380:499–505

    Article  PubMed Central  PubMed  Google Scholar 

  2. Brenner DJ, Hall EJ (2007) Computed tomography–an increasing source of radiation exposure. N Engl J Med 357:2277–2284

    Article  CAS  PubMed  Google Scholar 

  3. Smith AB, Dillon WP, Lau BC, Gould R, Verdun FR, Lopez EB et al (2008) Radiation dose reduction strategy for ct protocols: successful implementation in neuroradiology section. Radiology 247:499–506

    Article  PubMed  Google Scholar 

  4. Kulkarni NM, Uppot RN, Eisner BH, Sahani DV (2012) Radiation dose reduction at multidetector ct with adaptive statistical iterative reconstruction for evaluation of urolithiasis: how low can we go? Radiology 265:158–166

    Article  PubMed  Google Scholar 

  5. Vorona GA, Zuccoli G, Sutcavage T, Clayton BL, Ceschin RC, Panigrahy A (2013) The use of adaptive statistical iterative reconstruction in pediatric head ct: a feasibility study. Am J Neuroradiol 34:205–211

    Article  CAS  PubMed  Google Scholar 

  6. Maxfield MW, Schuster KM, McGillicuddy EA, Young CJ, Ghita M, Bokhari SA et al (2012) Impact of adaptive statistical iterative reconstruction on radiation dose in evaluation of trauma patients. J Trauma Acute Care Surg 73:1406–1411

    Article  PubMed Central  PubMed  Google Scholar 

  7. Kilic K, Erbas G, Guryildirim M, Arac M, Ilgit E, Coskun B (2011) Lowering the dose in head ct using adaptive statistical iterative reconstruction. AJNR Am J Neuroradiol 32:1578–1582

    Article  CAS  PubMed  Google Scholar 

  8. Rapalino O, Kamalian S, Kamalian S, Payabvash S, Souza LC, Zhang D et al (2012) Cranial ct with adaptive statistical iterative reconstruction: improved image quality with concomitant radiation dose reduction. AJNR Am J Neuroradiol 33:609–615

    Article  CAS  PubMed  Google Scholar 

  9. Hamberg LM, Rhea JT, Hunter GJ, Thrall JH (2003) Multi-detector row ct: radiation dose characteristics. Radiology 226:762–772

    Article  PubMed  Google Scholar 

  10. Wintermark M, Maeder P, Verdun FR, Thiran JP, Valley JF, Schnyder P et al (2000) Using 80 kvp versus 120 kvp in perfusion ct measurement of regional cerebral blood flow. Am J Neuroradiol 21:1881–1884

    CAS  PubMed  Google Scholar 

Download references

Acknowledgments

This project was supported by a research grant from GE Healthcare.

Conflict of interest

We declare that we have no conflict of interest.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Max Wintermark.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Table E1

(DOCX 24 kb)

Table E2

(DOCX 24 kb)

Table E3

(DOCX 24 kb)

Table E4

(DOCX 24 kb)

Table E5

(DOCX 24 kb)

Table E6

(DOCX 21 kb)

Table E7

(DOCX 21 kb)

Table E8

(DOCX 22 kb)

Table E9

(DOCX 22 kb)

Table E10

(DOCX 21 kb)

Table E11

(DOCX 22 kb)

Table E12

(DOCX 22 kb)

Table E13

(DOCX 21 kb)

Rights and permissions

Reprints and permissions

About this article

Cite this article

Komlosi, P., Zhang, Y., Leiva-Salinas, C. et al. Adaptive statistical iterative reconstruction reduces patient radiation dose in neuroradiology CT studies. Neuroradiology 56, 187–193 (2014). https://doi.org/10.1007/s00234-013-1313-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00234-013-1313-z

Keywords

Navigation